iSeg NHQ226L HV
iSeg NHQ226L HV
Set the polling rate. Note that if the polling is turned off,the information in the dialog may be invalid.
Voltage and Max Current settings
Data Formats
xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx
^^^^ ^^^^ ^^^^ ^^-----------------------data id
^^ ^^^^ ^^^^ ^^^^ ^^^^-length in longs
xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx
^^^^ ^^^^ ^^^^- device id
xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx time in seconds since Jan 1, 1970
xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx StatusWord1 for channel 0
xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx StatusWord2 for channel 0
xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx actual Voltage chan 0 encoded as a float
xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx actual Current chan 0 encoded as a float
xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx StatusWord1 for channel 1
xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx StatusWord2 for channel 1
xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx actual Voltage chan 1 encoded as a float
xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx actual Current chan 1 encoded as a float
The HV state. See the various symbols that can appear here and their meanings below.
The rate at which to ramp the voltage up or down
Stop ramping the voltage by loading in the last voltage read back
Load all settings to card. This may ramp the voltage to a new value if the set and act voltages differ
Panic the voltage to zero volts
Actual Voltage and Current as read back from card
Blank
V ==0
Ramping Down
Ramping Up
V<100
V>100
If there is a max current trip it will be indicated in this field
Note:
If the ‘DAQ’ switch on the front panel is NOT set to ‘DAQ’, the device is in manual mode and ORCA can not control it. Therefore, when in ‘Manual’ mode the dialog is deactivated