The SIS3800 is a 32-channel 32-bit scaler with a maximum counting frequency of 200 MHz (ECL and NIM) 100 MHz (TTL) respectively.


In ORCA it can be placed into either a standard VME crate or the VME64 crate.


See the manual for more specific information.

 

Read current scaler values, clear internal values, and continue. Doesn’t ship values.

Set the VME Address

How often to read the counters. Ships data unlessShipping only at End of Run’ option is selected

Data Production


In general, the scaler data records will be placed into the data stream at the polling frequency unless you have selected the ‘Ship ONLY at End of Run’ option. The manual options (i.e. pushing the ‘Read Now’ or ‘Read And Clear’ will NOT ship a data record. On the ‘Ship ONLY at End of Run’ option, the counters will be read at the end of run and the result will be shipped.



Waveform record:


xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx

^^^^ ^^^^ ^^^^ ^^----------------------- Data ID (from header)

-----------------^^ ^^^^ ^^^^ ^^^^ ^^^^- length

xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx

--------^-^^^--------------------------- Crate number

-------------^-^^^^--------------------- Card number

--------------------------------------^- 1==SIS38020, 0==SIS3000

xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx  time read in seconds since Jan 1, 1970

xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx  last time read in seconds since Jan 1, 1970 (zero if first sample)

xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx  count enabled mask

xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx  overFlow mask

xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx  options

-------------------------------------^^- lemo in mode

------------------------------------^--- enable25MHzPulses

-----------------------------------^---- enableInputTestMode

---------------------------------^------ enableReferencePulser

--------------------------------^------- clearOnRunStart

-------------------------------^-------- enable25MHzPulses

------------------------------^--------- syncWithRun

xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx  counts for chan 1

..

..

xxxx xxxx xxxx xxxx xxxx xxxx xxxx xxxx  counts for chan 32


Start/Stop the counting

Display either the raw scaler values or a live-time based on the count of the reference channel

You can edit  the channel names in place

Read current scaler values. Doesn’t ship values.

Load dialog settings to HW

Channel to use in the live-time calculations

Front panel lemo input modes and description

Reset to default settings. Stops counting.

On an overflow condition, the affected channel count values will turn red but will continue counting. You can clear the overflow condition here.